EE610 |
Low Differential Pressure Sensor |
|
The EE610 is dedicated for the highly accurate measurement of low differential pressure in medical, pharmaceutical and clean room applications. It can be employed in air or in non-flammable and non-aggressive gases. |
Type |
EE610 low differential pressure sensor for wallmounting with display |
|
EE610 low differential pressure sensor for wallmounting |
Measurement Performance |
The EE610 offers ± 0.5 Pa accuracy over the entire measurement range. For the versions with analogue outputs, the EE610 offers selectable measuring ranges of ± 25 / ± 50 / ± 100 Pa and 0…100 Pa. The long-term stable, piezo-resistive no-flow-through pressure sensing element excludes the risk of cross-contamination. |
Analogue and Digital Outputs |
The measured data is available on the analogue voltage and current outputs or on the RS485 interface with Modbus RTU or BACnet MS/TP protocol. |
Functional and Robust |
The IP65 / NEMA 4 enclosure facilitates easy mounting and minimizes installation costs. External mounting holes allow installation with closed cover. |
Configurable and Adjustable |
The setup and adjustment can be easily performed with DIP-switches and push buttons on the electronics board (EE610 with analogue outputs) or with an optional adapter and the free EE-PCS configuration software (EE610 with RS485 interface). The setup includes bidirectional or unidirectional pressure range, output signal, response time, display units and backlight. A zero and span point adjustment can be easily performed with push buttons on the electronics board. |
Key Features EE600 |
Multi-range device |
|
High accuracy |
|
Analogue or digital version available |
|
Fully configurable |
|
Large graphic display |
|
Installation-friendly enclosure |
Technical data EE600 |
|
MEASURING RANGE |
± 25 / ± 50 / ± 100 Pa |
|
0…100 Pa |
ACCURACY – *FULL SCALE (1000 PA OR 10000 PA) |
±0.5 Pa = 0.5 % FS* |
OVERLOAD LIMITS |
± 7000 Pa |
ANALOG OUTPUTS – SELECTABLE WITH DIP SWITCHES |
0-5 V or 0-10 V and |
|
0-20 mA or 4-20 mA |
DIGITAL OUTPUT |
Modbus RTU or BACnet MS/TP |