EE894

EE894 CO2 Module Measures Four Climate Parameters
  The digital module EE894 measures carbon dioxide (CO2), relative humidity (RH), temperature (T) and ambient pressure (p). Temperature and pressure compensation ensures high CO2 measurement accuracy under changing environmental conditions. Very small dimension and the I2C or E2 interface facilitate the design-in.
Detail The dual wavelength NDIR technology with auto-calibration is particularly long-term stable and insensitive to contamination. A multiple point CO2 and T factory adjustment leads to excellent CO2 measurement accuracy over the entire T working range. The pressure compensation minimizes the impact of altitude and ambient pressure variations onto the CO2 readings.
  The measured data, with a range of up to 1% CO2, is available on the I²C or the E2 digital interface. Small dimensions and electrical connection via contact pins, edge card socket or pads facilitate the design-in.
  An optional kit for the E2 interface facilitates easy configuration of the module and the adjustment of the CO2, RH, T and p measurement. The CO2 measurement interval can be set according to the application requirements; by this the average current consumption can be reduced to 420 μA, ideal for battery-operated devices.
Typical Applications Demand controlled ventilation
  Building automation
  Data loggers and hand helds
  Wireless transmitters
Key Features EE894 Dual wavelength NDIR CO2 working principle
  Measuring range up to 10,000ppm
  Auto-calibration for excellent long term stability
  Temperature and pressure compensation
  High insensitivity to pollution
  Easy user configuration and adjustment
  Very low current consumption
  Miniature size
Technical data EE894  
MEASURING RANGE CO2 0…2000 ppm
  0…5000 ppm
  0…10000 ppm
ACCURACY CO2 AT 25 °C (77 °F) AND 1013 MBAR (14.69 PSI) ±(50 ppm +2% of mv)
  ±(50 ppm +3% of mv)
  ±(100 ppm +5% of mv)
MEASURING RANGE RELATIVE HUMIDITY 0…95 % RH (non condensing)
ACCURACY RELATIVE HUMIDITY AT 25 °C (77 °C) AND 20…80% RH typ. ± 3 % RH
MEASURING RANGE PRESSURE 700…1100 mbar (10.15…15.95 psi)
ACCURACY PRESSURE AT 25 °C (77 °C) AND 20…80% RH typ. ± 2 mbar
MEASURING RANGE TEMPERATURE -40…60 °C (-40…140 °F)
ACCURACY TEMPERATURE AT 25 °C (77 °F) typ. ± 0.5 °C (± 0.9 °F)
DIGITAL INTERFACE I²C or E2
SUPPLY 4.75-7.5V DC